NIM Station for Ambient Gammadoserate

$ 1029.59

Availability: 22 in stock
  • Modification Description: Time Constant for Ratemeter 0.03 sec replaced with 600sec (big tank condenser added)
  • Condition: calibrated & long-run tested
  • Brand: BNC Portanim, Ortec Ratemeter, University, Hobo
  • Module Function: at present logging Ambient Gammadoserate
  • Modified Item: Yes

    Description

    - BNC Portanim Power-supply for NIM modules. The crate can take 4 units
    - Ortec 449-2 log-lin-Ratemeter, sends 10-Minute Countrate integral as DC Voltage to Hobo logger
    - Power Supply for GM-Tube, Pulses go to Ratemeter
    - Hobo Datalogger, receives data from ratemeter and can send data to smartphone
    - LND GM-Tube 7807
    energy compensated and capsuled , 12 m of Cable
    the Combination measures ambient gammadoserate with a big compensated gm-tube and sends the moving (gliding) average values (10min integration time) in 5-minute pace to the Hobo-datalogger.
    of course all parameters can be individually set by the customer.
    eg: the inherent bg of the gm tube is subtracted with the ratemeters precision dial at the bottom-center.
    alarms can be set all over the range
    one image shows the logged data of november-december last year. you see that heavy snowfall at
    26-27 november
    caused washout (snow-out...) of radon progeny: the airborne gamma-activity translated into the logged voltage went up to 1.74Volt, that translates to 1.74 x 62.5 nSv/h per Volt, equal 108,75 nSv/h or
    0.108𝞵Sv/h
    . (1 net-cps equals 62nanoSievert per hour. 1cps is represented at the back of the ratemeter by 1Volt. full range of the meter, 10cps, gives 10Volt at this BNC-socket)
    please contact me to find out freight cost to your location!